关键词:阈值;低能量起搏;程控
摘要 对117例安装不同品牌、型号起搏器的患者进行长期随访观察,探讨低能量起搏的安全性及程控计算方法。在安装起搏器2个月~12年后测定起搏阈值,结果为2.43±0.75 V、0.14±0.10 ms,同时将脉宽设置为阈值的2倍、3倍或更多。15例经1.1~4.3年的随访,再次重复测定阈值(8例增加、4例下降、3例不变),并调整相应参数。117例患者的随访时间为1~6年,此间均未发现起搏失效。由此认为起搏阈值设置为能量阈值的3~4倍是安全的。
Long-Term Observation in 117 Patients With Lower Energy Pacing
Zhao Jianrong, Zhou Zhihong, Qian Jianan
(Department of Cardiology,Ruijin Hospital, Affiliated Shanghai Second Medical
University, Shanghai,200025)
Abstract Various model pacemakers were implanted with 117 cases. The capture threshold was 2.43±0.75 V (pulse amplitude) and 0.14±0.10 ms (pulse width) after implanting for 2 months to 12 years. Then, The pulse width was set up two to three or more times of capture threshold. The capture threshold was measured repeatly after 1.1~ 4.3 years in 15 patients. No pacing failure was found in all patients in the following period. We recommed that a 3~ 4-fold margin of safety for capture is available based upon the deliveried pulse energy.
......
您现在查看是摘要页,全文长 15363 字符。