MIT单通道测量的电磁关系及脑组织电导率测量参数计算
http://www.100md.com 《第四军医大学学报》 2004年第21期
非接触测量,,脑组织;非接触测量;磁感应断层成像;电导率,0引言,1MIT实验系统与电磁关系模型,2基本理论假设和电磁关系求解,3模拟测量敏
     Electromagnetic relation of single measurement channel of MIT and computing of electromagnetic parameters in measurement of brain conductivity

    QIN MingXin, JIAO LiCheng, LI ShiJun, WANG Cong, L Hua, DONG XiuZhen

    1Research Center of Intelligent Information Processing, School of Electronic Engineering, Xidian University, Xi’an 710071, China, 2Department of Medical Electronic Engineering, School of Biomedical Engineering, Fourth Military Medical University, Xi’an 710033, China

    【Abstract】 AIM: To further optimize and improve upon our MIT experiment system and explore specific electromagnetic relations of our experiment system. METHODS: Based on electromagnetic induction theory and ideal physical model of our single measurement channel of MIT, we deduced electromagnetic relations of the system. RESULTS: Current phase and magnitude in the receiving coil were related to the conductivity of objects. The phase shifts and magnitude changes of the receiving coil current were calculated when the object conductivity was set for five sorts of normal brain tissues and brain edema. CONCLUSION: To improve the performance of MIT measurement system, we should use higher exciting frequency, accurately acquire phase shift between reference and receiving coils and choose phase sensitive detector of high performance. ......