睡眠障碍中心的认证标准(8)

     In circumstances of mixed use,defined as testing rooms being used for other medical testing or examination during nonsleep testing(daytime)hours,the testing room(s)must meet all of the space and equipment standards of a single use sleep testing room.

    Standard

    D6Testing BedroomsPhysical Characteristics

    All testing bedrooms must be single occupancy,private and comfortable,have hard walls,floortoceiling,and a privacy door that opens directly to a corridor or common use area such that the patient can access the testing bedroom without having to pass through another testing bedroom.

    Standard

    D7Testing Bedrooms & Emergency Care

    Patient testing bedrooms must not have any impediments to the delivery of emergency careThe patient testing rooms:

    amust be of sufficient size to accommodate emergency personnel access with a minimum of 24 inches of available clear space on 3 sides of the bed,bmust include a testing bed with a mattress not smaller than a standard hospital bed.

    Standard

    D8Bathroom Facilities

    The sleep facility must have clean bathrooms with a minimum ratio of one bathroom for every three testing rooms; these bathrooms must each contain a toilet and a sinkEach bathroom will have a working privacy doorSole access to a shared bathroom shall not be through a testing bedroom.

    Standard

    D9Handicap Testing Bedroom and Bathroom

    At least one testing bedroom and bathroom must be handicap accessible as defined either by local building regulations or section 63 and 64 of the American with Disabilities Act(http://wwwaccessboardgov/adaag/html/adaaghtm).

    Standard

    D10Control Room

    The dimensions of the control room must not be less than 40 square feet total or 20 square feet per testing bedroom,whichever is larger.

    Standard

    D11Communication

    The facility must maintain a twoway communication system between the patient bedroom and the control room and/or sleep facility personnel.

    Standard

    D12Video Recording

    Each testing bedroom in the facility must have a mechanism for visual monitoring and video recording of patients during testingTime delayed photographs will not be considered compliant with this standard.

    Standard

    D13Polygraphic Equipment

    The facility must maintain polygraphic equipment capable of recording and storing physiologic parameters using sensors,and recommended or alternative derivations as described in The AASM Manual for the Scoring of Sleep and Associated Events:Rules,Terminology,and Technical SpecificationsThe equipment must meet all of the “RECOMMENDED” minimal technical and digital specifications described in section III of the manual.